Class N, S and B are the three types of small steam sterilizers defined in the European standard EN 13060. Class N sterilizes only unwrapped solid instruments, Class S handles the loads specified by the manufacturer, and Class B uses pre-vacuum cycles to sterilize all loads, including wrapped, porous and hollow instruments.
Comparison table
| Class | Air removal | Suitable loads | Typical users |
|---|---|---|---|
| Class N | Gravity displacement | Unwrapped, solid instruments | Small clinics, basic labs |
| Class S | As specified by manufacturer (may include single vacuum) | Specified loads, some wrapped or hollow items | Clinics with mixed loads |
| Class B | Fractionated (pulsed) pre-vacuum | All loads: wrapped, porous, hollow and lumen instruments | Dental, surgical and hospital use |
Why air removal matters
Steam sterilizes only where it touches the surface. Air trapped inside wrapped packs, fabric or hollow tubes blocks steam, so those areas may not be sterilized. Class B machines pull a vacuum several times before heating, removing air completely. That is why a dental autoclave for handpieces is usually Class B.
Common sterilization cycles
- 121 °C at about 15 psi, typically 15–20 minutes hold time
- 134 °C at about 30 psi, typically 3–4 minutes hold time for unwrapped loads
Hold times depend on the load and should be validated.
What about large hospital and lab autoclaves?
Large sterilizers such as a horizontal autoclave or double door autoclave fall under large-sterilizer standards (such as EN 285), not EN 13060. For porous loads in hospitals, a pulse vacuum sterilizer does the same job as a Class B machine at larger scale. For media and lab waste, a vertical autoclave is often enough.
Frequently Asked Questions
Which autoclave class is best for dental clinics?
Class B is recommended for dental clinics because handpieces and wrapped instruments have hollow and porous parts that need pre-vacuum air removal.
Can a Class N autoclave sterilize wrapped instruments?
No. Class N is meant for unwrapped solid instruments only, because gravity air removal cannot clear air from wrapped or hollow loads.
How do I know if my autoclave cycle works?
Use chemical indicators, a Bowie-Dick test for pre-vacuum machines and biological indicators as per your validation schedule.