ICH Q1A(R2) sets the storage conditions for drug stability studies. For the general case, long-term testing is done at 25 °C ± 2 °C/60% RH ± 5% RH or 30 °C ± 2 °C/65% RH ± 5% RH, intermediate at 30 °C/65% RH and accelerated at 40 °C ± 2 °C/75% RH ± 5% RH.
General case storage conditions
| Study | Storage condition | Minimum period at submission |
|---|---|---|
| Long term | 25 °C ± 2 °C / 60% RH ± 5% RH or 30 °C ± 2 °C / 65% RH ± 5% RH | 12 months |
| Intermediate | 30 °C ± 2 °C / 65% RH ± 5% RH | 6 months |
| Accelerated | 40 °C ± 2 °C / 75% RH ± 5% RH | 6 months |
Refrigerated and frozen products
| Product | Long term | Accelerated |
|---|---|---|
| Stored in a refrigerator | 5 °C ± 3 °C | 25 °C ± 2 °C / 60% RH ± 5% RH |
| Stored in a freezer | -20 °C ± 5 °C | Case by case |
Climatic zones
Long-term conditions depend on where the product will be sold. The commonly used zones are Zone I (21 °C/45% RH), Zone II (25 °C/60% RH), Zone III (30 °C/35% RH), Zone IVa (30 °C/65% RH) and Zone IVb (30 °C/75% RH) for hot and very humid regions. Always confirm the conditions your regulator requires for each market.
Testing frequency
For long-term studies, samples are usually tested every 3 months in the first year, every 6 months in the second year and yearly after that. Accelerated studies need at least three time points, such as 0, 3 and 6 months.
Photostability (ICH Q1B)
Light testing exposes samples to an overall illumination of not less than 1.2 million lux hours and near-UV energy of not less than 200 watt hours per square metre, in a photo stability test chamber.
Which chambers do you need?
Most labs use one stability test chamber per condition so studies are not disturbed, a walk-in stability chamber for large volumes, a lab refrigerator for 5 °C studies and a deep freezer for -20 °C. Ask us for a stability chamber quotation.
Frequently Asked Questions
What are accelerated stability conditions?
Accelerated stability testing is done at 40 °C ± 2 °C/75% RH ± 5% RH for at least 6 months under ICH Q1A(R2).
What is intermediate stability testing?
Intermediate testing at 30 °C ± 2 °C/65% RH ± 5% RH is done when long-term testing is at 25 °C/60% RH and a significant change occurs during accelerated testing.
What is Zone IVb?
Zone IVb covers hot and very humid climates, with long-term stability conditions of 30 °C/75% RH.