ICH Stability Storage Conditions Explained (Q1A R2)

25 Nov, 2026
Admin User
2 min read
5M+ ProductsSold till date
Lab & IndustryTrusted equipment
Pan IndiaDelivery & support
FreeExpert consultation
In short We are manufacturer of Magnetic stirrer & Multiposition Magnetic stirrer with hot Plate in India. Call & Mail us for magnetic stirrer Prices

ICH Q1A(R2) sets the storage conditions for drug stability studies. For the general case, long-term testing is done at 25 °C ± 2 °C/60% RH ± 5% RH or 30 °C ± 2 °C/65% RH ± 5% RH, intermediate at 30 °C/65% RH and accelerated at 40 °C ± 2 °C/75% RH ± 5% RH.

General case storage conditions

StudyStorage conditionMinimum period at submission
Long term25 °C ± 2 °C / 60% RH ± 5% RH or 30 °C ± 2 °C / 65% RH ± 5% RH12 months
Intermediate30 °C ± 2 °C / 65% RH ± 5% RH6 months
Accelerated40 °C ± 2 °C / 75% RH ± 5% RH6 months

Refrigerated and frozen products

ProductLong termAccelerated
Stored in a refrigerator5 °C ± 3 °C25 °C ± 2 °C / 60% RH ± 5% RH
Stored in a freezer-20 °C ± 5 °CCase by case

Climatic zones

Long-term conditions depend on where the product will be sold. The commonly used zones are Zone I (21 °C/45% RH), Zone II (25 °C/60% RH), Zone III (30 °C/35% RH), Zone IVa (30 °C/65% RH) and Zone IVb (30 °C/75% RH) for hot and very humid regions. Always confirm the conditions your regulator requires for each market.

Testing frequency

For long-term studies, samples are usually tested every 3 months in the first year, every 6 months in the second year and yearly after that. Accelerated studies need at least three time points, such as 0, 3 and 6 months.

Photostability (ICH Q1B)

Light testing exposes samples to an overall illumination of not less than 1.2 million lux hours and near-UV energy of not less than 200 watt hours per square metre, in a photo stability test chamber.

Which chambers do you need?

Most labs use one stability test chamber per condition so studies are not disturbed, a walk-in stability chamber for large volumes, a lab refrigerator for 5 °C studies and a deep freezer for -20 °C. Ask us for a stability chamber quotation.

Frequently Asked Questions

What are accelerated stability conditions?

Accelerated stability testing is done at 40 °C ± 2 °C/75% RH ± 5% RH for at least 6 months under ICH Q1A(R2).

What is intermediate stability testing?

Intermediate testing at 30 °C ± 2 °C/65% RH ± 5% RH is done when long-term testing is at 25 °C/60% RH and a significant change occurs during accelerated testing.

What is Zone IVb?

Zone IVb covers hot and very humid climates, with long-term stability conditions of 30 °C/75% RH.

#ich stability conditions #ich q1a #stability storage conditions #climatic zones
Share:
A

Admin User

Yatherm Scientific manufactures and supplies laboratory, pharmaceutical, and industrial testing equipment. Our team writes these guides to help you choose and use the right instrument with confidence.

Keep Reading

You May Also Like

Looking for laboratory equipment?

Tell us your application and capacity. Our engineers will recommend the right model and send a quotation.